Applications
 

Acids, Aromatic and Plasticizers

Isomeric Aminobenzoic Acids, Phthalates - Silica

 

Operating Conditions

Sample Dissolved in: iso-octane: methylene 
chloride = 8:2
Adsorbent Used: DAI Silica 32-63
Stationary Phase: None
Eluent: methylene chloride
Sample Volume: 1.0 ml
Flow rate: 14 ml/min
Operating Pressure: 32 psi (for peaks 2&3)
Column Size: 9mm x 300 mm
Temperature: 20ºC
Detection: UV detector 253 nm

 

Substances Separated Sample (mg) Retention Volume (ml)
1. benzene 0.44 14.0
2. bis-(2-ethylhexyl)
phthalate (DOP)
0.50 77.0
3. diisobutyl phthalate 0.67 161.0
4. diethylphthalate (DEP) 0.67 350.0
Isomeric Aminobenzoic Acids, Phthalate - Silica

 

Operating Conditions

Sample Dissolved in: iso-octane: methylene 
chloride = 8:2
Adsorbent Used: DAI Silica 32-63
Stationary Phase: None
Eluent: methylene chloride
Sample Volume: 1.0 ml
Flow rate: 14 ml/min
Operating Pressure: 32 psi (for peaks 2&3)
Column Size: 9mm x 300 mm
Temperature: 20ºC
Detection: UV detector 253 nm
Substances Separated Sample (mg) Retention Volume (ml)
1. benzene 0.44 14.0 (16.0)*
2. bis-(2-ethylhexyl)
phthalate (DOP)
0.50 229.6 (51.2)
3. diisobutyl phthalate 0.67 364.0 (89.6)
4. diethylphthalate (DEP) 0.67 523.6 (140.8)