Applications

Analgesics, Pharmaceutical Formulations

Isopropylphenazone, Phenacetin - Silica

 

Operating Conditions

Sample Dissolved in: di-isopropyl ether
Adsorbent Used: DAI Silica 32-63
Stationary Phase: 20% water
Eluent: di-isopropyl ether: acetic acid = 98:2
Sample Volume: 0.15 ml
Flow rate: 20 ml/min
Operating Pressure: 35 psi
Column Size: 9mm x 300 mm
Temperature: 20ºC
Detection: UV detector 251 nm
Substances Separated Sample (mg) Retention Volume (ml)
1. benzene 4.0 10
2. isopropyl
phenazone
0.075 130
3. phenacetin 0.075 240

 